Electrochemical Work Station (Gamry-1)
Electrochemical Work Station (Gamry-1)
Make:
Model: Gamry 1010E (31194 & 31195)
Year of purchase: 2022
Funded by:
Location: Energy Lab
Contact email: energylab@cens.res.in
Book New Request

  • Potentiostat:
  • • 2- or 3- or 4-electrode configuration
  • • Floating (isolated from earth) or earth ground
  • • Maximum potential: ±12 V
  • • Maximum current: ± 1 A
  • • Compliance Voltage: ± 20 V
  • • Potentiostat rise time: < 1 µs
  • • Potentiostat bandwidth (-3 dB): >15 MHz
  • •Frequency range: 10 uHz – 2 MHz
  • •Maximum AC amplitude: 2.33 V rms
  • •Minimum AC amplitude: 17.8 uV rms
  • • Applied potential accuracy: ± 1 mV ± 0.2% of setting
  • • Applied potential resolution: 12.5 mV, 50 mV, 200 mV/bit
  • • Applied potential scan range: ± 0.4 V, ± 1.6 V, ± 6.4 V
  • • Measured potential accuracy: ± 1 mV ± 0.3% of setting
  • • Measured potential resolution: 400 mV, 100 mV, 10 mV, 1 mV/bit
  • • Applied currentaccuracy: ± 5 pA ± 0.3% of setting
  • • Applied currentresolution: 0.0033% full-scale/bit
  • • Measured current accuracy:± 5 pA ± 0.3% of setting
  • • Measured current resolution:0.0033% full-scale/bit
  • • Measured current bandwidth:> 10 MHz (100 mA - 100 mA ranges)
  • > 1.5 MHz (10 mA range)
  • > 150 kHz (1 mA range)
  • •iR compensation mode: Current interrupt
  • • Minimum interrupt time: 33 µs
  • • Maximum interrupt time: 715 s
  • Physical dimension:
  • • Weight: 2 kg
  • • Height: 24*6*27 cm (W*H*D)
  • • Cable: 60 cm
  • • Operational temperature: 0o C to 45o C
  • Electrometer:
  • • Reference electrode input impedance: 1012 ꭥ
  • • Reference electrode input bandwidth: > 15 MHz
  • • Reference electrode input bias current: < 10 pA
  •  Corrosion testing,
  •  Evaluation of protective coatings and paints,
  •  Research in electrochemical storage and conversion,
  •  Battery and electrochemical capacitor evaluation,
  •  Bio-electrochemical testing,
  •  Electrochemical sensor development.
  • • Chronoamperometry
  • • Chronopotentiometry
  • • Chronocoulometry
  • • Cyclic voltammetry
  • • Linear Sweep voltammetry
  • • Multistep chronoamperometry
  • • Multistep chronopotentiometry
  • • Potentiostatic EIS
  • • Galvanostatic EIS
  • • Hybrid EIS
  • • Cyclic charge discharge
  • • Self-discharge
  • • Open circuit potential measurement
  • • Tafel plot
  • • Differential pulse voltammetry
  • • Square wave voltammetry
  • • Sampled DC voltammetry
  • • Normal pulsed voltammetry
  • • Reverse normal pulsed voltammetry
  • • Chronopotentiometry (i-t curve for stability studies)
Modes Duration (hrs)
Sl. No. - Category Fees ( in Rs/- ) + additional 18% GST
01 - Academic Institutes 1,250 per slot / 8 hr
03 - External Industry Users 2,500 / 8 hr
02 - External Industry Partner 1,875 / 8 hr